Bought it used AS IS... figured can't go wrong for the price. Of course came with error E80 but was still usable and worked great after calibration. Replaced the battery and removed option01 (moved rear input back to the front). Still error E80. What? ...after troubleshooting I found that someone unsoldered the 100 ohm resistor in series with the battery. I don't know if this was done to render the unit useless or as part of a fully automatic computer controlled setup. Installed a new resistor and now have a fully functioning unit. Thank you.
Verified purchase: Yes | Condition: Pre-owned
Was sold as for repair or parts. Fortunately very little wrong with it. Also was in excellent cosmetic state. Problem with it was that the testing in the USA had been done on 110Volt while internal power key was set to 220v. So actually there was nothing of a repair needed and it worked just perfectly "out of the box". Lucky me. Thanks for great deal.
Verified purchase: Yes